TOP FLOW 2 | Print |  E-mail

Animal Change Station - Designed for emptying out and cleaning of cages, in a sterile area, with protection given to the operator and environment.

tn_300w_top-flow-2The Top Flow 2 Animal Change Station is a single fan cabinet manufactured like a Class I microbiological safety cabinet with inward air velocity more than 0.80 m/s, 100% exhausted through an HEPA filter and visual/acoustic alarms in compliance to European standard EN 12469: 2000.

This Animal Change Station is designed for emptying out and cleaning of cages, in a sterile area, with protection given to operator and environment. The unit is supplied with one garbage trolley with stainless steel wheels.

TOP FLOW 2 Technical Characteristics:

  • Exhaust HEPA filter: efficiency 99.95% MPPS H13 (99.99% 0.3 m DOP test), 305x610x292; with protection grid in order to avoid unattended blockage by foreign objects.
  • Pre filters: EU3 cells according to EN 779 400x500x22mm with quick fixing system and access through an hinged panel from the work area.
  • Prefilter protection grid made of stainless steel AISI 304L scotch brite finish.
  • Electronic board with microprocessor system to control all functioning of the cabinet: and automatic regulation of the fan speed fan to compensate the blockage of the filter, alarms, manage of the utilities.

Monitoring with alarms for the following parameters:

  1. Intake airflow velocity out of the pre-set range;
  2. Failure of the motor blower;
  3. Front window in wrong position;

Animal Change Station CleaningAnimal Change Station Waste DisposalAlphanumeric display showing:

  1. Laminar airflow speed in m/s;
  2. Date and time
  3. Hours counters for cabinet life
  4. UV Lamp run hours
  5. HEPA filters run hours
  6. Date of last HEPA filters change
  7. Date of last service
  8. Date of last power failure
  9. Change language
  10. Change password

Soft touch control panel.

  • Intake airflow velocity more than 0.8m/s* (as per Class I bio-safety cabinet) automatically adjusted and monitored by the microprocessor system. (*corresponding to 800 m3/h).
  • External structure made of steel epoxy polyester powder coated, resistant to the most common industrial disinfectant.
  • Hinged solid work surface made of stainless steel AISI 304L scotch brite finish supported by two gas springs, to allow access to the garbage bag to operator in standing position while the cabinet is running, with transition adapter to waste material into the garbage bag.
  • Hinged window in safety laminated glass framed with anodised aluminium profile supported by two gas springs.
  • Safety glass side windows.
  • Double inlet centrifugal fan with direct driven external rotor motor to improve efficiency end reduce noise (less than 58 dB).
  • Stainless steel swiveling wheels with brake.
  • Easy access to all the dirty zones inside the cabinet. 30-Watt fluorescent lamp in non-contaminated area.
  • Garbage bag trolley made of stainless steel AISI 304L scotch brite finish and plastic with swiveling wheels.

Norms and Directives:

As far as electrical safety is concerned, the cabinet is manufactured according to EN 61010 norm and in particular: EN 292-1, EN 292-2, EN 61010-1, EN 50081-1, EN 50082-1, EN 50081-2, EN 50082-2, IEC 801-2, IEC 801-4, ENV 50140, ENV 50141

Furthermore it has been designed to fulfill the European Directives like:

  • Machinery Directive 89/392/EEC, 91/368/EEC, 93/44/EEC, 93/68 EEC.
  • Electromagnetic Compatibility Directive 89/336/EEC, 92/31/EEC, 93/68/EEC.
  • The Masini Institute of Milan tested it and copy of the certification is available on request.
  • Low voltage Directive 72/23/EEC, 93/68/EEC.

IMPORTANT:

The cabinets we produce, according to the above statements, are manufactured in compliance with the laws on safety into force.
Particularly, they are manufactured in compliance with Art. 6 Dlgs 626/94 and its modification (Italian law in compliance with European Directive EEC/89/391) on the DUTY OF DESIGNERS, OF MANUFACTURERS, OF SUPPLIERS AND INSTALLATORS.
The cabinets are NOT submitted to the directive EEC 93/42 on Medical Devices, as indicated in the definition of "Medical Devices" reported in art. 1-point 2a of the Directive mentioned.
STERIL IS NOT responsible for damages to people and objects due to a non-proper use of the cabinet, and due to a fail to comply with the user instructions and maintenance that are always supplied with the cabinets.

TOP FLOW 2 Animal Change Station Characteristics:

Animal Change Station TOP FLOW 2
Technical Characteristics:
Overall dimension (WxDxH) 1100 x 840 x 1980 mm
Safe working area dimension (WxDxH)
1000 x 620 x 660 mm
Package dimension (WxDxH)
2100 x 980 x 1860 mm
Front opening dimension
290 mm
Max front opening height 630 mm
Net weight 200 Kg
System Ventilation Data:
Exhaust flow rate 870 m3/h
Heat emission at 25 °C (1) 600 Kcal/h
Noise level <55 dB (A)
Lighting
500 Lux
Electrical Data:
Voltage 230V AC F+N+P.E. V
Frequency 50 Hz
Power 700 W
Absorption 3 A
Electrical classification (with feeding cable) 1
Protection category IP 20
Fluorescent lamps
2x30 W


(1) Only for cabinets that recirculate air in environmental

Animal Change Station Side View Diagram Animal Change Station Front View Diagram

Animal Change Station Top View DiagramOverall Dimensions:

  1. Main Fan
  2. HEPA Filter
  3. Pre filters
  4. Charcoal filter (Option)
  5. Filter protection grid (removable)
  6. Slope hinged window with gas springs
  7. Funnel for waste material

 

Detail of soft touch control panel:

Animal Change Station Control Panel Diagram

The control panel accommodates:

  1. Digital display of the cabinet parameters
  2. Green LED for Power Voltage ON
  3. Esc key to cancel operation in progress or to exit from an alarm message
  4. ENTER button to input data
  5. Mode O (Cabinet OFF)
  6. Mode I (Ventilation ON)
  7. Mode II (UV + Service)
  8. Lighting ON/OFF push-button
  9. UV lamp ON/OFF push-button (option)
  10. Socket Outlet ON/OFF push-button (option)
  11. Fuel gas solenoid valve ON/OFF push-button
  12. RIGHT/LEFT arrow keys to scroll data
  13. UP and DOWN arrow keys to increase or decrease data (eg. time) 14) Red luminous indicator for alarm.
Last Updated on Monday, 23 March 2009 01:24